Product evaluation: failure analysis for fiber (b)(4): the fiber cap exhibits cratering and drilled through condition; the glass cap exhibits devitrification at the output area, and detritus adhesion around output area; the heat shrink tubing exhibits minor scratch marks.Based on device analysis, the potential for forward firing may exist.Probable root cause: based on the device analysis, the probable root cause of the failure is: heat accumulation.Cap wear was accelerated due to anatomical/procedural factors (tissue contact and technique) encountered during the procedure which would limit the performance of the fiber.
|
It was reported that during a bph surgical procedure at 3,693 joules and 25 minutes "fiber damaged at tip" and "didn't fire to the correct direction" were noted.The fiber was exchanged and the second fiber was used to complete the procedure at 298,570 joules.The tip of the fiber was cleaned during the procedure.No injury to the patient was reported.
|