MAUDE Adverse Event Report: ZOLL MEDICAL CORPORATION ASSY,MAIH,2.0 DIA, W/SWITCH, 100-120VONLY, PACKAGE; INTERNAL PADDLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MEDICAL CORPORATION ASSY,MAIH,2.0 DIA, W/SWITCH, 100-120VONLY, PACKAGE; INTERNAL PADDLE Back to Search Results
Model Number 1011-0139-05
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Information (3190)
Event Date 08/17/2015
Event Type  malfunction  
Manufacturer Narrative
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
 
Event Description
Complainant alleged that while attempting open chest defibrillation on a (b)(6) old male patient in ventricular fibrillation, the associated defibrillator displayed a "paddle error" message using these internal handles.Complainant indicated that the patient was placed on bypass and converted to an organized rhythm spontaneously.Complainant did not indicate that there was any adverse effect to the patient due to the reported malfunction.
 
Manufacturer Narrative
Zoll medical corporation evaluated the internal handles and the reported malfunction was not replicated or confirmed.The internal handles were put through extensive testing without duplicating the malfunction.Replacement internal handles were sent to the customer.No trend is associated with reports of this type.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASSY,MAIH,2.0 DIA, W/SWITCH, 100-120VONLY, PACKAGE
Type of Device
INTERNAL PADDLE
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
Manufacturer Contact
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key6855326
MDR Text Key85652418
Report Number1220908-2017-02151
Device Sequence Number1
Product Code LDD
UDI-Device Identifier00847946022365
UDI-Public00847946022365
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/21/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/11/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number1011-0139-05
Device Catalogue Number8011-0139-05
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/18/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received08/21/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age78 YR
-
-