• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NIHON KOHDEN CORPORATION ZM-541PA; TRANSMITTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NIHON KOHDEN CORPORATION ZM-541PA; TRANSMITTER Back to Search Results
Model Number ZM-541PA
Device Problems Incorrect Measurement (1383); Invalid Sensing (2293); Device Operational Issue (2914)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/15/2017
Event Type  malfunction  
Manufacturer Narrative
The biomedical engineer reported that the transmitter was giving inaccurate nibp readings.He tried new hoses and cuffs, as well as running it on a simulator, however the issue continued to persist.The device was in use on a patient, but no patient harm was reported.The biomedical engineer was provided with an exchanged transmitter.He sent the failed device in for repair and it is currently awaiting evaluation.Nihon kohden continues to investigate the reported event.
 
Event Description
The biomedical engineer reported that the transmitter was giving inaccurate nibp readings.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZM-541PA
Type of Device
TRANSMITTER
Manufacturer (Section D)
NIHON KOHDEN CORPORATION
1-31-4 nishiochia, shinjuku-ku
attn: shama mooman
tokyo, 161-8 560
JA  161-8560
Manufacturer (Section G)
NIHON KOHDEN CORPORATION
1-31-4 nishiochia, shinjuku-ku
attn: shama mooman
tokyo, gunma 161-8 560
JA   161-8560
Manufacturer Contact
shama mooman
safety mgmt dept, quality mgmt
seibu bldg 2, 4th floor 1-11-2
kusunokidai tokorozawa, saitama 359-8-580
JA   359-8580
MDR Report Key6860461
MDR Text Key86116258
Report Number8030229-2017-00301
Device Sequence Number1
Product Code DRT
UDI-Device Identifier04931921106617
UDI-Public04931921106617
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K043517
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 09/12/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/12/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberZM-541PA
Device Catalogue NumberZM-541PA
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/29/2017
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA09/12/2017
Distributor Facility Aware Date08/15/2017
Device Age23 MO
Event Location Hospital
Date Report to Manufacturer09/12/2017
Date Manufacturer Received09/12/2017
Was Device Evaluated by Manufacturer? No
Date Device Manufactured08/21/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-