|
Model Number DIMENSION VISTA 1500 |
Device Problem
Mechanical Problem (1384)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 08/21/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
The customer contacted the siemens customer care center (ccc).Quality control (qc) was within specification on the day the event occurred.The ccc specialist reviewed the instrument data which indicated that the integrated multi-sensor technology (imt) calibration was satisfactory but the peristaltic pump rate was low.The ccc specialist reviewed the process error log which showed imt probe pressure transducer error.A siemens customer service engineer (cse) was dispatched to the customer's site.After analyzing the instrument, the cse rebuilt the imt pump and found the imt metering pump full of air.The cse performed a total rebuild of the imt.The cse primed the system and ran quick check, resulting satisfactory.The cse reset the instrument and qc, which resulted within ranges.The cause of the discordant, falsely elevated na result is unknown.The instrument is performing within specifications.No further evaluation of the device is required.
|
|
Event Description
|
A discordant, falsely elevated sodium (na) result was obtained on a patient sample on a dimension vista 1500 instrument.The discordant result was not reported to the physician(s).The sample was auto repeated, resulting lower.The sample was then repeated on an alternate instrument, resulting the same as the auto repeated result.The alternate instrument result was reported to the physician(s).There are no reports of patient intervention or adverse health consequences due to the discordant, falsely elevated na result.
|
|
Search Alerts/Recalls
|
|
|