Catalog Number 82086-01 |
Device Problem
Tear, Rip or Hole in Device Packaging (2385)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 08/23/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Internal file number - (b)(4): during processing of this complaint, attempts were made to obtain complete event, patient and device information.The device was received.Investigation is not yet complete.A follow up report will be submitted with all relevant information.
|
|
Event Description
|
It was reported that before use, a hole was noted in the tyvek pouch of an xact stent delivery system from a 1 inch razor cut.The box itself did not have a razor cut.Another unspecified stent was used to successfully complete the procedure.There was no clinically significant delay in the procedure and no adverse patient effects.No additional information was provided.
|
|
Manufacturer Narrative
|
(b)(4).Visual inspections were performed on the returned device.The reported packaging damage was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history revealed no other incidents.The investigation was unable to determine a cause for damaged packaging.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|
|