• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MITG-OKLAHOMA CITY TRUCLEAR; INSUFFLATOR, HYSTEROSCOPIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MITG-OKLAHOMA CITY TRUCLEAR; INSUFFLATOR, HYSTEROSCOPIC Back to Search Results
Model Number 7210165-
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/21/2017
Event Type  malfunction  
Manufacturer Narrative
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during set up last (b)(6), the device had scratch on the lens.There was no patient involvement.
 
Manufacturer Narrative
Evaluation summary post market vigilance (pmv) led an evaluation of one device.A visual inspection was performed on the product and aspiration spigot is broken off.Complaint of broken spigot was confirmed.Spigot has never been replaced and is over 4 years old.Other than the damage to spigot the mdu passes functional testing including high speed testing (100-2500 rpms) with blade installed.Unit passed functional tests on a known good control unit with footswitch.The complaint investigation has concluded that this unit has succumbed to damage to spigot.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.The root cause of the observed condition was determined to be a result of a misuse of the product.Damage may occur if the headpiece is handled roughly during use.Excessive forces may result in traumatic insertion, chipping particles of the optical glass, or other damage to the unit.Should new information become available, the file will be re-opened and the investigation summary will be amended as appropriate.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUCLEAR
Type of Device
INSUFFLATOR, HYSTEROSCOPIC
Manufacturer (Section D)
MITG-OKLAHOMA CITY
75 s. meridianave
oklahoma OK
Manufacturer (Section G)
MITG-OKLAHOMA CITY
75 s. meridianave
oklahoma OK
Manufacturer Contact
sharon murphy
60 middletown ave
north haven, CT 06473
2034925267
MDR Report Key6871921
MDR Text Key86454545
Report Number1643264-2017-21090
Device Sequence Number1
Product Code HIG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K031616
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/18/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number7210165-
Device Catalogue Number7210165-
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/05/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received12/22/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-