Catalog Number 9733457
Device Problem Bent (1059)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/23/2017
Event Type  malfunction  
Manufacturer Narrative
Patient information not provided due to japanese patient privacy regulations.Device lot number, or serial number, unavailable.) device manufacturing date is dependent on lot number/serial number, therefore, unavailable.No parts have been received by manufacturer for evaluation.Part not returned for evaluation.
 
Event Description
A medtronic representative reported that during a spinal fusion procedure the tip of the cervical probe was bent.Site used another probe to complete the procedure.The procedure was completed with the use of navigation.There was no delay to the procedure.No impact on patient outcome.
 
Manufacturer Narrative
Additional information: device lot number and manufacturer date provided.The suspect probe was returned to the manufacturer for evaluation.Testing found that the probe was slightly bent and that it passed functional testing.The hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.25 MM PEDICLE PROBE
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
peter verhey
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
7208902515
MDR Report Key6872920
MDR Text Key86492581
Report Number1723170-2017-03796
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00613994335968
UDI-Public00613994335968
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K050438
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
Report Date 12/06/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/18/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Catalogue Number9733457
Device Lot Number160405
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/10/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received11/10/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/05/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-