• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WAVELIGHT GMBH WAVELIGHT EX500 EXCIMER LASER; OPHTHALMIC EXCIMER LASER SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WAVELIGHT GMBH WAVELIGHT EX500 EXCIMER LASER; OPHTHALMIC EXCIMER LASER SYSTEM Back to Search Results
Catalog Number 8065990794
Device Problem Output above Specifications (1432)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Event Description
A service technician reported high voltage at (b)(4) percent of refractive treatment.Additional information has been requested.
 
Manufacturer Narrative
Additional information provided.The device history records (dhr) for the device was reviewed.No abnormalities that could have contributed to this event were found.The associated device was released based on company acceptance criteria.At the visit on site the field service engineer replaced both 45° deflector and focus lenses to fix this issue.Visual inspection of both the 45° deflectors and focus lenses shows burnings and scratches on all parts.The root cause for the reported problem is the wearing on the 45° deflectors and focus lenses through the usage over long time in the field.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WAVELIGHT EX500 EXCIMER LASER
Type of Device
OPHTHALMIC EXCIMER LASER SYSTEM
Manufacturer (Section D)
WAVELIGHT GMBH
am wolfsmantel 5
erlangen 91058
GM  91058
Manufacturer (Section G)
WAVELIGHT GMBH
am wolfsmantel 5
erlangen 91058
GM   91058
Manufacturer Contact
nadia bailey
am wolfsmantel 5
erlangen 91058
GM   91058
8176152330
MDR Report Key6878999
MDR Text Key87865730
Report Number3003288808-2017-02030
Device Sequence Number1
Product Code LZS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P020050
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/20/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065990794
Other Device ID Number00380659907944
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/20/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received02/16/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/16/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-