A siemens technical business lead (tbl) remotely reviewed the error log and found no issue with the integrated multi-sensor technology (imt) system.The tbl also noticed that the customer had recently loaded fresh quality control (qc) solution and ran it from a cup and from a vial.The customer had also replaced the standard a solution, after which quick check on imt module, was run and passed.The customer aligned the imt and imt probe.A siemens customer service engineer was dispatched to the customer site.While inspecting the instrument, the cse found microbubbles in the a line.The cse replaced the pump tubing and consumables.The cse reran quality control (qc), resulting within acceptable range.The cause of the discordant, falsely elevated na result is unknown.The instrument is performing according to specifications.No further evaluation of the device is required.
|
A discordant, falsely elevated sodium (na) result was obtained on a patient sample on a dimension vista 500 instrument.The discordant result was reported to the physician(s) who questioned it.The sample was repeated on an alternate dimension vista instrument, resulting lower and as expected.The alternate instrument result was reported to the physician(s).There are no reports of patient intervention or adverse health consequences due to the discordant, falsely elevated na result.
|