BOSTON SCIENTIFIC - GALWAY WATCHMAN® ACCESS SYSTEM; SYSTEM, APPENDAGE CLOSURE, LEFT ATRIAL
|
Back to Search Results |
|
Model Number M635TS10060 |
Device Problem
Leak/Splash (1354)
|
Patient Problem
Blood Loss (2597)
|
Event Date 08/23/2017 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).Device evaluated by mfr: returned product consisted of the watchman access sheath (was) with dilator inside the sheath as received.The valve was opened as received and blood was on the device.The valve, hub, shaft, and tip were microscopically and tactile inspected.Inspection revealed numerous kinks in the sheath, with the threads of the was were damaged/cross threaded.It could not be determined when the thread damage occurred.Inspection of the inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.Functional testing of the valve confirmed the ability to completely close the valve with forward force.Water was injected into the was by attaching a syringe filled with water.There were no leaks or air bubbles observed during water injection.The root cause is not confirmed as there was no evidence of the alleged issue or any anomalies which could have contributed to the reported difficulty.(b)(4).
|
|
Event Description
|
It was reported a valve leak occurred.A left atrial appendage (laa) closure procedure was being performed.A watchman® access system (was) was used.During the procedure, the hemostasis valve was found to be leaking blood after the valve was fully closed.The patient experienced a drop in hemoglobin.The valve was tightened onto the dilator and it was turning, but it did form a seal.The leaking was attempted to be stopped by thumb pressure, but it was not sufficient.The procedure was completed with this access system.The patient¿s condition was stable.
|
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
It was initially reported that the valve was tightened onto the dilator and it was turning, but it did form a seal.It was further clarified that the valve was not tightened onto the dilator.It was turning, but it did not form a seal.
|
|
Search Alerts/Recalls
|
|
|