The device history record (dhr) review confirmed that the device met all material, assembly and performance specifications.The subject device was returned wrapped around itself.The guidewire used with the catheter was not returned.Visual inspection revealed that the balloon catheter was compressed at approximately 18.0cm distal to the strain relief.It was also kinked at approximately 4.5cm from its distal end.The inflation port and proximal lumen were inspected under microscopy and the inflation port shaft appeared to be perforated.The inner shaft under the manifold appeared to be perforated.There was crystallized substance (appeared to be saline) found in the catheter as well.There were no issues observed with the coating and no other anomalies were observed.During functional testing, the 0.014" demonstration guidewire and a.010" mandrel was introduced and advanced through the catheter proximal end.The guidewire and mandrel could not be passed through the catheter, a lot of friction was encountered.The balloon failed to inflate with air because of the perforation/holes in the balloon and water leaked out of the balloon during the inflation testing.The balloon could not stay inflated.The kinks and compressions found on the catheter during device analysis were determined to be a result of handling damage during the clinical procedure.Information available indicated that the device was confirmed to be in good condition prior to use and the air leak was observed during aspiration.It is likely that the found catheter shaft leak was interpreted by the user as a balloon leak.Per the device directions for use (dfu): "introduce the guidewire, flexible-end first, into the straight (back) port of the manifold.To avoid kinking, advance the guidewire slowly in small increments to the end of the balloon catheter.¿ based on the examination of the damage inside the inflation lumen (balloon port), it appears that the guidewire was inserted into the balloon port (instead of the guidewire lumen), leading to the observed and reported leak.Based on the information available, an assignable cause of use error was assigned to this event.
|