Results: the penumbra system ace 68 reperfusion catheter (ace68) was kinked approximately 68.0 cm from the hub.Conclusions: evaluation of the returned device revealed that the ace68 was kinked.This type of damage typically occurs due to improper handling during preparation.If the device is handled with force during preparation, damage such as this may occur.During functional analysis, a mandrel was advanced through the hub of the ace68; however, the mandrel could not be advanced beyond the kink in the ace68.Penumbra catheters are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
While preparing a penumbra system ace 68 hi-flow kit (kit) for a thrombectomy procedure, the radiologic technologist noticed that the distal tip of the penumbra system ace 68 reperfusion catheter (ace68) was crimped.The damaged ace68 was found prior to use in the patient and therefore, the ace68 was not used for the procedure.The procedure was completed using a new kit.
|