• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. MODULAR FEMORAL STEM PRESS-FIT PLASMA SPRAYED CEMENTLESS SIZE 15; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. MODULAR FEMORAL STEM PRESS-FIT PLASMA SPRAYED CEMENTLESS SIZE 15; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Information (3190)
Event Date 08/25/2017
Event Type  Injury  
Manufacturer Narrative
(b)(4).Multiple mdr's were submitted for this event.Please see reports: 0001822565-2017-06570, 0001822565-2017-06571.(b)(4).(b)(6).The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Not returned to manufacturer.
 
Event Description
It was reported that foreign debris was found inside the sterile barrier.Surgery was delayed for 31-60 minutes.No further information has been made available at this time.
 
Manufacturer Narrative
Complaint sample was evaluated and the reported event was confirmed.The stem and it's packaging was returned and from returned product evaluation it was determined that received ldpe bag (87089971800) exhibits black substance on the inside of the bag in the area of the plasma spray coated portion of the stem.Dhr was reviewed and no discrepancies were found.Review of the complaint history determined that no further action is required.The root cause can be design deficiency and transit damage.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MODULAR FEMORAL STEM PRESS-FIT PLASMA SPRAYED CEMENTLESS SIZE 15
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key6885997
MDR Text Key87789720
Report Number0001822565-2017-06570
Device Sequence Number1
Product Code LWJ
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
PK0633251
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/22/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberN/A
Device Catalogue Number00771301500
Device Lot Number62815410
Other Device ID NumberSEE H10
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/28/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/26/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/30/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Other;
-
-