MAUDE Adverse Event Report: COVIDIEN MFG DC BOULDER LIGASURE; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN MFG DC BOULDER LIGASURE; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number LF1537
Device Problem Device Or Device Fragments Location Unknown (2590)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/04/2017
Event Type  Injury  
Manufacturer Narrative
Evaluation summary: one used lf1537 ligasure device was returned, and examination of the sample identified an issue with a detached component.Visual inspection found the metal flange on the device jaw had broken off and was not returned.The tube is made of stainless steel and should not break without exerting excessive force or abuse.The damage to the metal flange is consistent with the device being processed or used multiple times.The investigation identified the root cause of the reported event to be user error.The instructions for use included with this device cautions users that the device is only intended for single use.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
The customer reported that the device broke during use on a patient.No patient injury occurred or medical intervention was required.Examination of the received device by medtronic found the metal flange on the lower jaw had broken off and was not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGASURE
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
Manufacturer (Section G)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
Manufacturer Contact
sharon murphy
5920 longbow drive
boulder, CO 80301
2034925267
MDR Report Key6889286
MDR Text Key87166054
Report Number1717344-2017-05950
Device Sequence Number1
Product Code GEI
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K092879
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Health Professional
Type of Report Initial
Report Date 08/31/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/24/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberLF1537
Device Catalogue NumberLF1537
Device Lot Number50570023X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/21/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/31/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age57 YR
-
-