Evaluation summary: one used lf1537 ligasure device was returned, and examination of the sample identified an issue with a detached component.Visual inspection found the metal flange on the device jaw had broken off and was not returned.The tube is made of stainless steel and should not break without exerting excessive force or abuse.The damage to the metal flange is consistent with the device being processed or used multiple times.The investigation identified the root cause of the reported event to be user error.The instructions for use included with this device cautions users that the device is only intended for single use.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|