• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC NEUROMODULATION PRECISION; SPINAL CORD STIMULATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC NEUROMODULATION PRECISION; SPINAL CORD STIMULATOR Back to Search Results
Model Number SC-1110-02
Device Problems Failure to Charge (1085); High impedance (1291); Telemetry Discrepancy (1629)
Patient Problems Inadequate Pain Relief (2388); No Known Impact Or Consequence To Patient (2692)
Event Date 06/15/2017
Event Type  malfunction  
Manufacturer Narrative
Additional suspect medical device component involved in the event: model#: sc-8216-70 serial #: (b)(4) description: artisan surgical lead, 70cm.
 
Event Description
A report was received that the patient was not able to charge the ipg and lead was showing high impedance reading.It was noted that the patient was having inadequate stimulation and the patient¿s remote control (rc) kept saying communication failed.The patient will undergo a lead and ipg replacement procedure.Device malfunction was suspected on the ipg.
 
Manufacturer Narrative
Additional information was received that no further course of action will be taken at this time.It is indicated that the devices will not be returned for evaluation; therefore a failure analysis of the complaint devices could not be completed.A review of the device history records will be conducted.If there is any further relevant information from that review, a supplemental med watch will be filed.
 
Event Description
A report was received that the patient was not able to charge the ipg and lead was showing high impedance reading.It was noted that the patient was having inadequate stimulation and the patient¿s remote control (rc) kept saying communication failed.The patient will undergo a lead and ipg replacement procedure.Device malfunction was suspected on the ipg.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRECISION
Type of Device
SPINAL CORD STIMULATOR
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
Manufacturer Contact
talar tahmasian
25155 rye canyon loop
valencia, CA 91355
6619494863
MDR Report Key6891161
MDR Text Key87463023
Report Number3006630150-2017-03605
Device Sequence Number1
Product Code LGW
UDI-Device Identifier08714729767688
UDI-Public08714729767688
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
030017
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/15/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/25/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date01/07/2017
Device Model NumberSC-1110-02
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/12/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/20/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age70 YR
-
-