COOK ENDOSCOPY 6 SHOOTER SAEED MULTI-BAND LIGATOR; MND, LIGATOR, ESOPHAGEAL Back to Search Results
Catalog Number MBL-U-6
Device Problem Detachment Of Device Component (1104)
Patient Problem Foreign Body In Patient (2687)
Event Type  Injury  
Manufacturer Narrative
Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.The device history record could not be reviewed because a lot number was not provided.Investigation conclusion: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.The outer diameter of the endoscope used during the procedure is unknown.The user is advised to use an endoscope with an outer diameter between 9.5 mm - 13 mm.Use of an endoscope with an outer diameter outside the recommended range can result in barrel detaching from the endoscope.The instructions for use advise the user: "refer to package label for minimum channel size required for this device." the instructions for use state under system preparation: "attach barrel to tip of endoscope, ensuring barrel is advanced onto tip as far as possible." the instructions for use direct the user to: "lubricate endoscope and exterior portion of barrel." the barrel can dislodge if lubricant is applied to the endoscope prior to attaching the barrel or if lubricant is allowed inside the barrel before the loading process.The instructions for use state: "caution: do not place lubricant inside barrel." a possible contributing factor to barrel detachment includes allowing body fluids to enter the area between the barrel and endoscope.The instructions for use states under precautions: "prior to assembling device, routine endoscopic examination is recommended to confirm diagnosis requiring treatment of esophageal varices or internal hemorrhoids." if the endoscope is wiped free after the initial check for location of the varices, body fluids in this area can be avoided.Prior to distribution, all 6 shooter saeed multi-band ligators are subjected to a visual inspection to ensure device integrity.Corrective action: a review of the complaint history was conducted.The likelihood of occurrence is considered rare.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Event Description
During an endoscopic variceal ligation (evl), the physician used a cook 6 shooter saeed multi-band ligator.As reported to customer relations, "the head of the product (which houses the ligator) [barrel] became detached from the strings [barrel detachment].The detached part was retrieved from the patient's stomach using a net.A new ligator was attached and the procedure was completed.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
6 SHOOTER SAEED MULTI-BAND LIGATOR
Type of Device
MND, LIGATOR, ESOPHAGEAL
Manufacturer (Section D)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer Contact
scottie fariole
4900 bethania station rd
winston-salem, NC 27105
3367440157
MDR Report Key6896867
MDR Text Key87465317
Report Number1037905-2017-00608
Device Sequence Number1
Product Code MND
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial
Report Date 09/27/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/27/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberMBL-U-6
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Event Location Hospital
Date Manufacturer Received09/06/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
OLYMPUS ENDOSCOPE - UNKNOWN MODEL
Patient Outcome(s) Required Intervention;
-
-