MEDTRONIC PUERTO RICO OPERATIONS CO. 640G INSULIN PUMP MMT-1711K; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
|
Back to Search Results |
|
Model Number MMT-1711K |
Device Problem
Erratic or Intermittent Display (1182)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 09/21/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.The insulin pump involved in this event is the 640g insulin infusion pump, which is not marketed in the united states.However, the device is similar to the paradigm real-time insulin infusion pump, which is marketed in the united states.
|
|
Event Description
|
It was reported that, the pump flashing display.The blood glucose level was 270 mg/dl at the time of incident.The pump has been dropped during customer was playing outside.The insulin pump will not be returned for analysis.
|
|
Manufacturer Narrative
|
After battery installation unit gave an unexpected flashing white / blank display followed by an unexpected intermittent beep alarm due to vertical cracked lcd controller.Unable to perform the self test, sleep current measurement, active current measurement, displacement test or verify missing segments/partial display due to display anomaly.
|
|