Model Number E602 |
Device Problem
High Test Results (2457)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 05/24/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).This event occurred in (b)(6).
|
|
Event Description
|
The customer had issues with questionable elecsys testosterone ii assay results for proficiency material and patient samples.Data was provided for three patient samples.Patient 1 initial result was 8.47 nmol/l and the repeat result was 2.01 nmol/l.Patient 2 initial result was 6.31 nmol/l and the repeat result was 0.64 nmol/l.Patient 3 initial result was 5.06 nmol/l and the repeat result was 0.602 nmol/l.Only the erroneous initial result for patient 2 was reported outside the laboratory.Only the repeat results were reported for patients 1 and 3.There was no allegation of an adverse event.The reagent lot number and expiration date were requested but were not provided.Multiple service visits were performed and it was determined the extra wash station (ews) sipper lines and the reagent probe were blocked.These were cleaned.It was also found that the procell syringe was contaminated and had crystallization.This was drained, removed, and cleaned.Signs of liquid and dirt were found on the valve, which was cleaned.The measuring cells were changed and new tubing, nozzles, and probes were installed.Analyzer performance testing was completed and was acceptable.
|
|
Manufacturer Narrative
|
The field service representative performed photomultiplier high voltage adjustment, a blank cell calibration, and calibration, qc, and precision testing.The customer stated they were still having issue with erroneous testosterone results for qc material.
|
|
Manufacturer Narrative
|
The field service representative checked the liquid level detection (lld) and the probe adjustments.The probe horizontal sampling position was adjusted and the syringe unit was replaced.Performance testing was completed and was acceptable.
|
|
Manufacturer Narrative
|
The degasser was replaced.The system was decontaminated, and no further issues were reported.A definitive root cause could not be determined.
|
|
Search Alerts/Recalls
|