Age at time of event: 18 years or older.Device evaluated by manufacturer: the returned product consisted of the watchman access sheath (was).The valve was opened as received and blood inside and outside of the device.The valve, hub, shaft, and tip were microscopically and tactile inspected.Inspection of the threads of the was revealed damaged (cross threaded).Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.Functional testing of the valve confirmed the ability to completely close the valve with forward force.Water was injected into the was by attaching a syringe filled with water.There were no leaks or air bubbles observed during water injection.The root cause is not confirmed as there was no evidence of the alleged issue or any anomalies which could have contributed to the reported difficulty.(b)(4).
|