The customer experienced an ongoing issue with questionable ise indirect gen.2 sodium results.Data was provided for two patient samples.Patient 1 initial result was 146 mmol/l with a data flag and the repeat result was 140 mmol/l.Patient 2 initial result was 119 mmol/l and the repeat result was 140 mmol/l.The initial results were reported outside of the laboratory.The repeat results were believed to be correct.The patients were not adversely affected.The electrode lot number and expiration date were requested but were not provided.The customer received calibration alarms, replaced all the ise reagents and performed an ise check which was acceptable.The customer ran qc which was acceptable.The field service representative found there was a misadjusted sample probe and rinse nozzle which he adjusted.He checked the fluidic and pneumatic systems and ran calibration, qc, and precision testing with no errors and all values within specification.
|