MAUDE Adverse Event Report: DATASCOPE FAIRFIELD SENSATION PLUS 8FR. 50CC IAB; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE FAIRFIELD SENSATION PLUS 8FR. 50CC IAB; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Catalog Number 0684-00-0575
Device Problems Bent (1059); Sticking (1597); Failure to Advance (2524)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/10/2017
Event Type  malfunction  
Manufacturer Narrative
The product was returned with the membrane completely unfolded and blood on the exterior of the catheter with the one-way valve attached.The guide wire returned was found to be bent near the middle and could not be used for testing.One kink was found on the catheter tubing near the y-fitting approximately 75.4cm from the iab tip.The technician attempted to insert a laboratory 0.025¿ guide wire through the inner lumen and was able to successfully insert the guide wire.No obstructions were felt.The evaluation confirmed the reported problem because the guide wire was found to be bent.We are unable to determine when this may have occurred.A device and lot history record review was completed for the reported product.No non-conformances were found that are considered to be related to the event.(b)(4).
 
Event Description
It was reported that during insertion the intra-aortic balloon (iab) was stuck on a bent guidewire and was unable to be advanced.There was no injury or harm to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSATION PLUS 8FR. 50CC IAB
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE FAIRFIELD
15 law drive
fairfield NJ 07004
Manufacturer (Section G)
DATASCOPE FAIRFIELD
15 law drive
fairfield NJ 07004
Manufacturer Contact
15 law drive
fairfield, NJ 07004
MDR Report Key6914925
MDR Text Key89698253
Report Number2248146-2017-00437
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112327
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Nurse
Type of Report Initial
Report Date 10/04/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/05/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Expiration Date04/10/2020
Device Catalogue Number0684-00-0575
Device Lot Number3000049199
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/15/2017
Is the Reporter a Health Professional? Yes
Device AgeYR
Date Manufacturer Received09/11/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/10/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age75 YR
Patient Weight78
-
-