Catalog Number BI70002000 |
Device Problems
Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138); Device Operates Differently Than Expected (2913)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 09/11/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Patient information was unavailable from the site.No parts were replaced.No parts have been received by manufacturer for evaluation.
|
|
Event Description
|
A site representative reported that during a electrode and probe placement the imaging system froze and needed reboot.The procedure was completed with the use of imaging.There was no delay to procedure.No impact on patient outcome.
|
|
Manufacturer Narrative
|
A medtronic representative went to the site to test the equipment.The representative reported that a washer on the x-ray tube was replaced to resolve the reported issue.The imaging system then passed the system checkout and was found to be fully functional.
|
|