• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PLEXUS MANUFACTURING SDN. BHD; PULSE-GENERATOR, PACEMAKER, EXTERNAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PLEXUS MANUFACTURING SDN. BHD; PULSE-GENERATOR, PACEMAKER, EXTERNAL Back to Search Results
Model Number 5392
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Product event summary: analysis was unable to confirm the customer comment that the device needed the new battery shorting bar design, this update had been performed previously.It was noted that there was evidence of a person not affiliated with the manufacturer disassembling and reassembling the device.The main seal was installed incorrectly, the liquid crystal display (lcd) flex was damaged, the tube sleeve was missing, all case and printed circuit board screws were torqued incorrectly, the red and white wires on the lcd were routed incorrectly as were the battery wires.Contamination was found throughout the device: the lower case, hanger assembly, display flex, multiple screws, keypad flex, encoder assembly and nuts and encoder knobs.The white wire on the printed circuit board (pcb) connector of the output connector was not fully seeded into the connector anymore, the main pcb was corroded and all six plugs were damaged.Due to the excessive contamination the device was to be scrapped.This device was reported as included in the field action noted but returned product investigation found the device did not perform as described in the field action.The device is no longer included as part of the field action.
 
Event Description
The external pulse generator was returned per instructions for an advisory and subsequently tested out of specification during manufacturer's analysis.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
PULSE-GENERATOR, PACEMAKER, EXTERNAL
Manufacturer (Section D)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY  11900
Manufacturer (Section G)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY   11900
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key6919620
MDR Text Key88446679
Report Number3004593495-2017-00497
Device Sequence Number1
Product Code DTE
UDI-Device Identifier00643169508767
UDI-Public00643169508767
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K150246
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Remedial Action Recall
Type of Report Initial
Report Date 07/11/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/06/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number5392
Device Catalogue Number5392
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/05/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received07/11/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/13/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberZ-1356-2015
Patient Sequence Number1
-
-