Product event summary: analysis was unable to confirm the customer comment that the device needed the new battery shorting bar design, this update had been performed previously.It was noted that there was evidence of a person not affiliated with the manufacturer disassembling and reassembling the device.The main seal was installed incorrectly, the liquid crystal display (lcd) flex was damaged, the tube sleeve was missing, all case and printed circuit board screws were torqued incorrectly, the red and white wires on the lcd were routed incorrectly as were the battery wires.Contamination was found throughout the device: the lower case, hanger assembly, display flex, multiple screws, keypad flex, encoder assembly and nuts and encoder knobs.The white wire on the printed circuit board (pcb) connector of the output connector was not fully seeded into the connector anymore, the main pcb was corroded and all six plugs were damaged.Due to the excessive contamination the device was to be scrapped.This device was reported as included in the field action noted but returned product investigation found the device did not perform as described in the field action.The device is no longer included as part of the field action.
|