MAUDE Adverse Event Report: MEDTRONIC IRELAND MICRA; LEADLESS PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND MICRA; LEADLESS PACEMAKER Back to Search Results
Model Number MC1VR01
Device Problem Battery Problem (2885)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/15/2017
Event Type  malfunction  
Event Description
It was reported that prior to being implanted, the battery longevity of the leadless implantable pulse generator (ipg) was below the limit for implantation.The device was opened and not used.Another device was used.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: performance data collected from the device was received and analyzed.Analysis of device memory indicated an issue with battery depletion out of box.Battery voltage was 2.958 volts prior to implant.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Product event summary: the device was returned and analyzed.The returned device indicated that the battery voltage was below implant specification.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICRA
Type of Device
LEADLESS PACEMAKER
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key6920811
MDR Text Key89509866
Report Number9612164-2017-01385
Device Sequence Number1
Product Code PNJ
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P150033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 06/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/06/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Medical Equipment Company Technician/Representative
Device Expiration Date06/28/2018
Device Model NumberMC1VR01
Device Catalogue NumberMC1VR01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/30/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/06/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/12/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-