Model Number MC1VR01 |
Device Problem
Battery Problem (2885)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 08/15/2017 |
Event Type
malfunction
|
Event Description
|
It was reported that prior to being implanted, the battery longevity of the leadless implantable pulse generator (ipg) was below the limit for implantation.The device was opened and not used.Another device was used.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Product event summary: performance data collected from the device was received and analyzed.Analysis of device memory indicated an issue with battery depletion out of box.Battery voltage was 2.958 volts prior to implant.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Product event summary: the device was returned and analyzed.The returned device indicated that the battery voltage was below implant specification.If information is provided in the future, a supplemental report will be issued.
|
|