MPRI SELECTSECURE; DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIAL (RA) PACEMAKER ELE Back to Search Results
Model Number 383059
Device Problems High impedance (1291); Over-Sensing (1438); Electronic Property Issue (2928)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/22/2017
Event Type  Injury  
Event Description
It was reported that during use the right atrial lead exhibited oversensing, that was replicated with pocket manipulation.It was also reported that the rv lead exhibited high impedance with polarity switch to unipolar.The rv lead was explanted and replaced.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the full lead was returned and analyzed.No anomalies were found.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during use the right atrial (ra) lead exhibited oversensing, that was replicated with pocket manipulation.It was also reported that the ra lead exhibited high impedance with polarity switch to unipolar.The ra lead was explanted and replaced.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SELECTSECURE
Type of Device
DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIAL (RA) PACEMAKER ELE
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key6924770
MDR Text Key88791440
Report Number2649622-2017-12155
Device Sequence Number1
Product Code NVN
Combination Product (y/n)N
Reporter Country CodeNL
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/07/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/06/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date05/08/2017
Device Model Number383059
Device Catalogue Number383059
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/04/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/13/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/28/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age35 YR
-
-