Model Number 176630
Device Problems Retraction Problem (1536); Mechanics Altered (2984)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/12/2017
Event Type  malfunction  
Event Description
According to the reporter, after 5 or 6 firings were successful, surgeon could not load the device anymore.No patient injury.
 
Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The instrument was received partially fired with five clips remaining.Visual inspection of the instrument noted that one of the jaw legs was out of position.The cam on the jaw leg was not aligned with the driver.Functionally, the jaw leg was reset by aligning the cam with the driver.The instrument was then applied to test media.The remaining clips loaded into the jaws, formed properly, released from the jaws and remained securely attached to test media.The interlock engaged after all clips were dispensed to prevent the jaws from approximating.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures; therefore, a device history record will not be performed.Replication of the disengaged jaw cam condition may occur when one of the jaws legs gets forcefully pulled outward away from the center line of the shaft.The noted jaw cam disengagement impedes functionality of the jaws, possibly resulting in clip malformation and/or difficulty removing the instrument from a trocar.The root cause of the observed damage was misuse of the product which would have caused or contributed to the reported incident.Should new information become available, the file will be re-opened and the investigation summary will be amended as appropriate.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO CLIP III
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DO  0101
Manufacturer (Section G)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DO   0101
Manufacturer Contact
sharon murphy
60 middletown ave
north haven, CT 06473
2034925267
MDR Report Key6930407
MDR Text Key89475073
Report Number9612501-2017-06366
Device Sequence Number1
Product Code FZP
UDI-Device Identifier20884521057859
UDI-Public20884521057859
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K071406
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/07/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/09/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2020
Device Model Number176630
Device Catalogue Number176630
Device Lot NumberJ7C1534X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/06/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/01/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/20/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial