Model Number 0292 |
Device Problems
Device Contamination With Biological Material (2908); Device Dislodged or Dislocated (2923)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 09/01/2017 |
Event Type
Injury
|
Manufacturer Narrative
|
At this time, the product has not been returned.If the product is returned, analysis will be performed and this report would be updated at that time.
|
|
Event Description
|
Boston scientific received information that this right ventricular (rv) lead dislodged.However, right before the procedure, the right atrial (ra) thresholds were also observed to have increased, so both leads were scheduled to be revised.During the procedure, the physician was unable to reposition the rv lead and extend the helix.The patient started coughing up blood, so the physician opted to explant and replaced both leads as there was concern about a pneumothorax.However, additional information confirmed there was no pneumothorax.Both leads were successfully implanted with no additional adverse patient effects.
|
|
Manufacturer Narrative
|
Upon receipt at our post market quality assurance laboratory, a thorough evaluation of the lead was performed.Resistance and pressure tests were completed to assess lead electrical performance and insulation integrity.Measurements throughout these tests were within normal limits.The lead did not pass the helix mechanism test, but was most likely due to body fluid infiltration.After the blood/body fluid was loosened, the helix was functional.The dislodgement was not confirmed as the lead tip and helix had no visible signs of damage or defect which would lead to dislodgement.Helix mechanism was confirmed, so the observation of helix difficulty was likely due to blood/body fluid infiltration into the helix mechanism.
|
|
Search Alerts/Recalls
|
|