The valve was implanted to the patient via lp-shunt.Doi was unknown and with initial setting 4.After a few day, ventricular enlargement was observed.The pressure setting was reduced but the issue continued.Obstruction was suspected and reoperation was performed on (b)(6) during the procedure, it was found that the valve and the sg were ruptured.Therefore, the valve was explanted and implanted a new one.The product will be returned to your site.
|
The valve was returned for evaluation.The position of the cam when valve was received was at setting 1.The valve was visually inspected: the housing was torn at the level of the siphonguard.The valve was tested for programming according and passed.The valve was flushed, no occlusion was noted.The valve was not reflux and leak tested as the housing was torn.The siphon guard was removed.No issues were found with the siphonguard.The valve was then pressure tested and passed the test.Review of the history device records was not possible as the lot number is unknown.The root cause of the occlusion suspected by the customer was not identified as no occlusion was found during the investigation.The root cause for the tear/cut in the silicone housing is probably due to the user.As noted in the ifu silicone has a low tear / cut resistance.Through a previous investigation, it was concluded that silicone housing tears are not design related, in order for the housing tear to occur the user has to compromise the silicone through a nick or tear in order for the event to occur.Validation testing demonstrates a robust design.Testing has shown that if the silicone housing has been compromised, a housing tear is likely to occur.Trends will be monitored for this or similar complaints.At present, we consider this complaint to be closed.
|