• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - SPENCER RX CYTOLOGY BRUSH; ENDOSCOPIC CYTOLOGY BRUSH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - SPENCER RX CYTOLOGY BRUSH; ENDOSCOPIC CYTOLOGY BRUSH Back to Search Results
Model Number M00545000
Device Problem Detachment Of Device Component (1104)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/19/2017
Event Type  malfunction  
Manufacturer Narrative
The complainant was unable to provide the suspect device lot number; therefore, the lot expiration and device manufacture dates are unknown.However, the complainant reported that the device was not expired.Device code relates to problem code for the reported event of radiopaque marker detached.Although the suspect device has been received, the evaluation has not been completed.Therefore, the cause of the reported malfunction has not been determined.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
 
Event Description
It was reported to boston scientific corporation that an rx cytology brush was used in the bile duct during a scraping cytology procedure performed on (b)(6) 2017.According to the complainant, during the procedure, the radiopaque marker of the brush detached and was left inside the bile duct.The radiopaque marker was left to passed naturally.The procedure was completed with this device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be fine.
 
Manufacturer Narrative
Investigation results: visual analysis of the returned rx cytology brush found that the ro marker was not present and was not returned for evaluation.It was noted that there were indentation marks found at the ro marker section which indicates proper assembly during the manufacturing process.The complaint that the radiopaque marker detached was confirmed.Due to anatomical/procedural factors encountered during the procedure; performance of the device was limited.Therefore, the most probable root cause classification is operational context.
 
Event Description
It was reported to boston scientific corporation that an rx cytology brush was used in the bile duct during a scraping cytology procedure performed on (b)(6) 2017.According to the complainant, during the procedure, the radiopaque marker of the brush detached and was left inside the bile duct.The radiopaque marker was left to passed naturally.The procedure was completed with this device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RX CYTOLOGY BRUSH
Type of Device
ENDOSCOPIC CYTOLOGY BRUSH
Manufacturer (Section D)
BOSTON SCIENTIFIC - SPENCER
780 brookside drive
spencer IN 47460
Manufacturer (Section G)
BOSTON SCIENTIFIC - SPENCER
780 brookside drive
spencer IN 47460
Manufacturer Contact
nancy cutino
100 boston scientific way
marlborough, MA 01752
5086834000
MDR Report Key6936881
MDR Text Key90057244
Report Number3005099803-2017-03012
Device Sequence Number1
Product Code FDX
UDI-Device Identifier08714729268628
UDI-Public08714729268628
Combination Product (y/n)N
PMA/PMN Number
K930348
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/19/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/10/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM00545000
Device Catalogue Number4500
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/22/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/13/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-