The customer's quality control (qc) was within range on the day the event occurred.A siemens customer service engineer (cse) was dispatched to the customer site.The cse replaced the integrated multi-sensor technology (imt) rotary valve assembly, peristaltic pump assembly, and imt tubings.The cse ran a system check, which passed.The cse ran qc for electrolytes, and level 3 chloride (cl) was out low.The cse replaced all fluids and sensor.The cse verified the sample probe alignments and conditioned the system.The cse ran dilution check.The cse reran qc, resulting within specification.The customer requested that a preventative maintenance (pm) be performed.The cause of the discordant, low na results is unknown.The instrument is performing according to specifications.No further evaluation of the device is required.
|
The customer obtained discordant low sodium (na) results on a dimension xpand plus with hm instrument.The samples were repeated and still resulted low and were reported to the physician(s), who questioned them.The samples were repeated again on the same instrument, resulting normal as expected.The corrected result was reported to the physician(s).There are no known reports of patient intervention or adverse health consequences due to the discordant low na results.
|