AV-TEMECULA-CT HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL II GUIDE WIRE Back to Search Results
Catalog Number 1009664
Device Problems Peeled/Delaminated (1454); Difficult To Position (1467); Difficult to Remove (1528)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/22/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device is expected to be returned for evaluation.It has not yet been received.A follow up report will be submitted with all relevant information.
 
Event Description
It was reported that during the percutaneous coronary intervention (pci) procedure, the bmw universal ii guide wire was advanced to the lesion.The intravascular ultrasound (ivus) catheter was advanced over the guide wire however resistance was noted.The devices were removed as a whole system.After removal from the anatomy, the ivus catheter was removed with resistance noted.The black coating of the guide wire was noted to be peeled off.Another guide wire was used to complete the procedure.There was no clinically significant delay in the procedure and no adverse patient effects.No additional information was provided.
 
Manufacturer Narrative
(b)(4).The device was returned for analysis.The reported peeled polymer and difficulty removing the device were unable to be confirmed.The reported difficulty to position was able to be confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents from this lot.It is possible that as the guide wire was advanced to the lesion interaction with the anatomy and/or manipulation of the device resulted in the noted stretched tip coils thus as the intravascular ultrasound catheter was advanced over the guide wire resulted in the reported difficult to position and the reported difficulty to remove, however this cannot be confirmed.The investigation determined a conclusive cause for the reported/noted difficulties cannot be determined.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL II GUIDE WIRE
Type of Device
GUIDE WIRE
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
BARCELONETA, PUERTO RICO REG# 3005737652
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key6940468
MDR Text Key90294788
Report Number2024168-2017-08136
Device Sequence Number1
Product Code DQX
Combination Product (y/n)N
Reporter Country CodeHK
PMA/PMN Number
K072460
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 11/07/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/11/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2019
Device Catalogue Number1009664
Device Lot Number7032371
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/26/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/06/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/01/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-