Catalog Number 1009664 |
Device Problems
Peeled/Delaminated (1454); Difficult To Position (1467); Difficult to Remove (1528)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 09/22/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device is expected to be returned for evaluation.It has not yet been received.A follow up report will be submitted with all relevant information.
|
|
Event Description
|
It was reported that during the percutaneous coronary intervention (pci) procedure, the bmw universal ii guide wire was advanced to the lesion.The intravascular ultrasound (ivus) catheter was advanced over the guide wire however resistance was noted.The devices were removed as a whole system.After removal from the anatomy, the ivus catheter was removed with resistance noted.The black coating of the guide wire was noted to be peeled off.Another guide wire was used to complete the procedure.There was no clinically significant delay in the procedure and no adverse patient effects.No additional information was provided.
|
|
Manufacturer Narrative
|
(b)(4).The device was returned for analysis.The reported peeled polymer and difficulty removing the device were unable to be confirmed.The reported difficulty to position was able to be confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents from this lot.It is possible that as the guide wire was advanced to the lesion interaction with the anatomy and/or manipulation of the device resulted in the noted stretched tip coils thus as the intravascular ultrasound catheter was advanced over the guide wire resulted in the reported difficult to position and the reported difficulty to remove, however this cannot be confirmed.The investigation determined a conclusive cause for the reported/noted difficulties cannot be determined.There is no indication of a product quality issue with respect to manufacture, design or labeling.
|
|
Search Alerts/Recalls
|