(b)(6).Complaint sample was evaluated and the reported event was confirmed.Product evaluation stated, "spherical radius and spherical radius depth were measured and found to be conforming.Features measured 2.2mm and 2.25mm respectively.The instrument shows typical signs of wear and tear.Dhr was reviewed and no discrepancies were found.Review of the complaint history determined that no further action is required as no were trends identified.Investigation results concluded that the most likely cause of ink removal is extended wear or delamination of the ink from the substrate due to autoclave cycling.This report is being submitted late as it has been identified in remediation.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|