The circumferential cautery probe was placed in the patient through enteroscopy scope and several cautery applications were done.The probe was removed and the tip was wiped down per procedure.The probe was then place in the scope again.When the tip was visualized, the tip of the probe was gone.The probe was removed from the field and given to clinical engineering.When questioned further, the author of the internal report stated that the surgeon was sure the missing portion of the tip detached when the instrument was being wiped down, and that it had not fallen in the patient.There was no harm to the patient and the procedure was completed as planned.
|