MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC. FORTIFY ASSURA VR ICD_IDE_US; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC. FORTIFY ASSURA VR ICD_IDE_US; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD1257-40Q
Device Problems Premature End-of-Life Indicator (1480); Battery Problem (2885)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/30/2017
Event Type  Injury  
Event Description
Following the battery performance alert (bpa) advisory, a bpa was received by the clinician and the device was explanted.The asymptomatic patient was stable post procedure.
 
Event Description
New information noted that elective replacement indicator and end of life indicator were both triggered when the patient arrived for the implantable cardioverter defibrillator replacement.
 
Manufacturer Narrative
Premature battery depletion was confirmed by analysis.No sources of high current were noted.The cause of the premature battery depletion was consistent with li cluster formation.From these analyses, in the absence of high current draw, it is probable that the premature battery depletion was caused by a lithium cluster induced short circuit.Li clusters are a known depletion mechanism for these advisory products that has been investigated and associated with a field action in (b)(6) 2016.The device is included in the premature battery depletion with implantable cardioverter defibrillator advisory notice issued by st.Jude medical on (b)(6) 2016.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FORTIFY ASSURA VR ICD_IDE_US
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key6949428
MDR Text Key89340564
Report Number2938836-2017-32872
Device Sequence Number1
Product Code NVZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P910023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Nurse
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date 05/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/13/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2014
Device Model NumberCD1257-40Q
Device Lot Number4110684
Other Device ID Number05414734505475
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/12/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/21/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/07/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-0115- 2017
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age59 YR
-
-