Catalog Number 8732-7101 |
Device Problem
Disassembly (1168)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 09/20/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Without a product return, no product evaluation is able to be conducted.The lot number is unknown; therefore the device history records are unable to be reviewed.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.
|
|
Event Description
|
It was reported that an attachment pin fell out of a slap hammer during surgery.The pin fell into and was removed from the wound.An alternative instrument was not needed to complete the procedure.There were no reports of patient impacts associated with this event.
|
|
Manufacturer Narrative
|
Current information is insufficient to permit a valid conclusion about the cause of this event.A follow up report will be sent upon completion of the device evaluation.
|
|
Manufacturer Narrative
|
Additional information: the returned slap hammer was evaluated.One of the quick connect attachment pins was found to have disassembled from the remainder of the device.The cause cannot be conclusively determined; however possible contributors include high forces placed on the device while trying to remove a too large trial, rasp, or spacer from the disc space or repetitive cleanings/washings.A review of the manufacturing records did not identify any issues which would have contributed to this event.
|
|