• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET SPINE INC. SLAP HAMMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET SPINE INC. SLAP HAMMER Back to Search Results
Catalog Number 8732-7101
Device Problem Disassembly (1168)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/20/2017
Event Type  malfunction  
Manufacturer Narrative
Without a product return, no product evaluation is able to be conducted.The lot number is unknown; therefore the device history records are unable to be reviewed.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.
 
Event Description
It was reported that an attachment pin fell out of a slap hammer during surgery.The pin fell into and was removed from the wound.An alternative instrument was not needed to complete the procedure.There were no reports of patient impacts associated with this event.
 
Manufacturer Narrative
Current information is insufficient to permit a valid conclusion about the cause of this event.A follow up report will be sent upon completion of the device evaluation.
 
Manufacturer Narrative
Additional information: the returned slap hammer was evaluated.One of the quick connect attachment pins was found to have disassembled from the remainder of the device.The cause cannot be conclusively determined; however possible contributors include high forces placed on the device while trying to remove a too large trial, rasp, or spacer from the disc space or repetitive cleanings/washings.A review of the manufacturing records did not identify any issues which would have contributed to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SLAP HAMMER
Type of Device
HAMMER
Manufacturer (Section D)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
westminster CO 80021
Manufacturer (Section G)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
na
westminster CO 80021
Manufacturer Contact
geoffrey gannon
10225 westmoor dr.
na
westminster, CO 80021
3034437500
MDR Report Key6949984
MDR Text Key90181145
Report Number3012447612-2017-00558
Device Sequence Number1
Product Code FZY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PEXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 03/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/13/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8732-7101
Device Lot NumberTU01095
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/12/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/22/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/24/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-