Model Number N/A |
Device Problem
Difficult to Open or Close (2921)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The complaint device has been returned, but the device investigation has not yet been completed.Once the evaluation is completed, a supplemental medwatch 3500a will be submitted.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
It was reported that foam was attached to top of packaging, which when opened, kicked the liner in its package to the ground.No adverse events have been reported as a result of the malfunction.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was confirmed.Inspection of the returned device revealed that the inner foam is adhered to the tyvek lid as reported thus confirming the reported event.Evidence of adhesive transfer is present around the full length and width of the cavity flange.Device history record was reviewed and no discrepancies were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further event information available at the time of this report.
|
|
Manufacturer Narrative
|
This follow-up report is being filled to relay additional information, which was unknown a the time of the initial medwatch.Udi # - (b)(4).
|
|