• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. ARCOM XL 44-36 STD HMRL BRNG; PROSTHESIS, SHOULDER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. ARCOM XL 44-36 STD HMRL BRNG; PROSTHESIS, SHOULDER Back to Search Results
Model Number N/A
Device Problems Difficult to Insert (1316); Device Damaged Prior to Use (2284)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/19/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that the surgeon was assembling the poly to the tray on the back table when the surgeon noticed a lip on the poly that was preventing it from seating correctly.The surgeon asked for a new poly to be opened, and the case went on successfully.No adverse events have been reported as a result of the malfunction.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated.Complaint sample was evaluated and the reported event was confirmed.Humeral bearing received for evaluation.As received the bearing exhibited damaged lip.It is reported by sales representative that assembly tool was not used during the procedure.Dimensions taken are within specification.Device history record was reviewed and no discrepancies were found.Review of the complaint history determined that no further action is required as no trends were identified.Operational context contributed to the event as the humeral bearing assembly tool was not used during the procedure.A summary of the investigation has been sent to the complainant.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCOM XL 44-36 STD HMRL BRNG
Type of Device
PROSTHESIS, SHOULDER
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key6954053
MDR Text Key90534879
Report Number0001825034-2017-08637
Device Sequence Number1
Product Code PAO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK080642
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/22/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/17/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date08/14/2022
Device Model NumberN/A
Device Catalogue NumberXL-115363
Device Lot Number511050
Other Device ID NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/21/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/21/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/14/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Age75 YR
-
-