This is a supplemental report to provide additional information based on the evaluation of the device.The subject device was returned to olympus medical systems corp.(omsc).Omsc investigated it on november 2, 2017.The needle of the subject device could not be retracted into the outer tube.The outer tube of the subject device was kinked at approximately 270 mm, 870 mm, 1030 mm, and 1420 mm from the distal end.The manufacturing record was reviewed and found no irregularities.The reported event is most likely related to the operator's technique.Omsc surmised that the kinks of the outer tube occurred due to applying bending load to the outer tube when the tube was inserted into the endoscope, the subject device was unpacked from the sterile bag, or pre-use inspection was performed.The frictional resistance between the outer tube and the needle tube increased due to the kinks, and then the needle could not be retracted into the outer tube.The instruction manual of the device has already warned as follows; warnings: insert the instrument slowly.Abrupt insertion could damage the endoscope and/or instrument.Stop using the instrument if the insertion portion bends excessively during use.This could result in malfunction, such as failing to extend the needle or inject a fluid.
|