MAUDE Adverse Event Report: TRUMPF MEDIZIN SYSTEME GMBH + CO. KG MARS OPERATING TABLE; SURGICAL TABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TRUMPF MEDIZIN SYSTEME GMBH + CO. KG MARS OPERATING TABLE; SURGICAL TABLE Back to Search Results
Model Number 1429733
Device Problems Unintended System Motion (1430); Device Issue (2379); Unintended Movement (3026)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/18/2017
Event Type  malfunction  
Manufacturer Narrative
The remote control was exchanged and investigated.Visible heat related damage was present at the housing and membrane keypad as a result of impact with a heat generating device.Subsequent electrical testing identified an electrical problem was caused by the heat related damage.After replacing the membrane keypad the remote control and operating table worked as intended.It is unknown what caused the heat damage.
 
Event Description
A trumpf medical mars operating table began to move after plugging in the cable remote control.The leg sections started moving downwards.No patient impact reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MARS OPERATING TABLE
Type of Device
SURGICAL TABLE
Manufacturer (Section D)
TRUMPF MEDIZIN SYSTEME GMBH + CO. KG
carl-zeiss strasse 7-9
saalfeld, 07318
GM  07318
Manufacturer Contact
steffen ulbrich
carl-zeiss strasse 7-9
saalfeld, 07318
GM   07318
MDR Report Key6954203
MDR Text Key90561014
Report Number3007143268-2017-00022
Device Sequence Number1
Product Code JEA
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Reporter Occupation Other
Remedial Action Inspection
Type of Report Initial
Report Date 09/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/17/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1429733
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/18/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/11/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-