• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC. TENDRIL STS; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC. TENDRIL STS; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 2088TC/52
Device Problem Difficult To Position (1467)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/25/2017
Event Type  malfunction  
Event Description
During the implant procedure, the right atrial (ra) lead could not be positioned due to the lead body rotating while extending the helix.The lead was replaced and the same event was noted on the second ra lead.The second ra lead was also removed and replaced, and the patient was in stable condition.Related manufacturer report number: 2017865-2017-32853.
 
Manufacturer Narrative
As received, a complete lead was returned in one piece with the helix extended and clogged with tissue.Visual and x-ray imaging revealed the outer insulation was twisted due to over-torque of the inner coil at the connector region.After cleaning, the helix retracted and extended.The over-torqued inner coil contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TENDRIL STS
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
15900 valley view court
sylmar CA 91342
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key6955362
MDR Text Key89935512
Report Number2017865-2017-32852
Device Sequence Number1
Product Code NVN
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P960013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/06/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/17/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2020
Device Model Number2088TC/52
Device Lot NumberP000048244
Other Device ID Number05414734502887
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/05/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/19/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/02/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-