During the implant procedure, the right atrial (ra) lead could not be positioned due to the lead body rotating while extending the helix.The lead was replaced and the same event was noted on the second ra lead.The second ra lead was also removed and replaced, and the patient was in stable condition.Related manufacturer report number: 2017865-2017-32853.
|
As received, a complete lead was returned in one piece with the helix extended and clogged with tissue.Visual and x-ray imaging revealed the outer insulation was twisted due to over-torque of the inner coil at the connector region.After cleaning, the helix retracted and extended.The over-torqued inner coil contributed to the event.
|