MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION SARNS TEMP. CONTROL AND MONITOR SYSTEM (TCM II); CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS-SARNS TCM II HEATING AND COOLING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION SARNS TEMP. CONTROL AND MONITOR SYSTEM (TCM II); CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS-SARNS TCM II HEATING AND COOLING Back to Search Results
Model Number 164925
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/22/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).As per field service representative, a 'pump not primed' error message was displayed and the pump not prime light was illuminated.There were no issues with heating or cooling and no leaks were observed.The customer chose not to have the unit serviced.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.(b)(4).
 
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, there was black debris in both the large and small tank.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SARNS TEMP. CONTROL AND MONITOR SYSTEM (TCM II)
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS-SARNS TCM II HEATING AND COOLING
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
Manufacturer Contact
katie hoyt
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key6955653
MDR Text Key90444784
Report Number1828100-2017-00475
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K883603
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial
Report Date 10/17/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/17/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number164925
Device Catalogue Number164925
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received09/25/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/17/1990
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-