• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MITG - OKLAHOMA CITY TRUCLEAR; HYSTEROSCOPE (AND ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MITG - OKLAHOMA CITY TRUCLEAR; HYSTEROSCOPE (AND ACCESSORIES) Back to Search Results
Model Number 7209808-
Device Problems Break (1069); Detachment Of Device Component (1104)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/13/2017
Event Type  malfunction  
Manufacturer Narrative
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, the device¿s hand piece broken off during testing.There was no patient involvement.
 
Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The complaint investigation has concluded that this unit has succumbed to damage to mdu receptacle.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.The root cause of the observed condition was determined to be a result of a misuse of the product.The ifu recommends "careful inspection of the operative hysteroscope before and after the procedure for possible signs of damage.Immediate detection and repair of minor damage will extend the life of the operative hysteroscope.¿ damage may occur if the user attempts to connect a damaged or defective hand piece to the control unit.Should new information become available, the file will be re-opened and the investigation summary will be amended as appropriate.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUCLEAR
Type of Device
HYSTEROSCOPE (AND ACCESSORIES)
Manufacturer (Section D)
MITG - OKLAHOMA CITY
75 s. meridian ave
oklahoma city OK 73107
Manufacturer (Section G)
MITG - OKLAHOMA CITY
75 s. meridian ave
oklahoma city OK 73107
Manufacturer Contact
sharon murphy
60 middletown ave
north haven, CT 06473
2034925267
MDR Report Key6959892
MDR Text Key90449350
Report Number1643264-2017-21121
Device Sequence Number1
Product Code HIH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K132015
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/18/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number7209808-
Device Catalogue Number7209808-
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/20/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received01/04/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-