• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PERCUSSIONAIRE CORP. PERCUSSIONAIRE; IPV HC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PERCUSSIONAIRE CORP. PERCUSSIONAIRE; IPV HC Back to Search Results
Model Number F00012-HT
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/24/2017
Event Type  malfunction  
Manufacturer Narrative
The ipv hc is a homecare therapy device for airway clearance.The suspect device was returned to the distributor's biomed technician, who completed the investigation.The device was investigated and functional tested with no issues noted, and subsequently returned into service.During the technician's investigation, no evidence of heat or smoke was found on or in the device.The only anomolies noted were that the gray outer layer of the power cord insulation was shifted, and exposed a bit more of the inner wires' insulation than typical, though the insulation on the inner wires was intact, with no copper wires showing.Also, the fan grills were missing, and the filter cap was missing.This was not a reportable event, but we decided to file this mdr so that our response to the end user's report would be on record.
 
Event Description
Home health nurse reported that the device was hot and smoking, and that the wires on the unit's connector were frayed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERCUSSIONAIRE
Type of Device
IPV HC
Manufacturer (Section D)
PERCUSSIONAIRE CORP.
1655 glengary bay rd.
sagle ID 83860
Manufacturer (Section G)
PERCUSSIONAIRE CORP.
1655 glengary bay rd.
sagle ID 83860
Manufacturer Contact
shawn burns
1655 glengary bay rd.
sagle, ID 83860
2082632549
MDR Report Key6962415
MDR Text Key90753345
Report Number1000524541-2017-00004
Device Sequence Number1
Product Code NHJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K905236
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Reporter Occupation Other
Remedial Action Other
Type of Report Initial
Report Date 10/02/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/19/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Home Health Aide
Device Model NumberF00012-HT
Device Catalogue NumberS00009
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received08/16/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/07/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age8 YR
Patient Weight29
-
-