• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - CORK ENCORE BALLOON CATHETER INFLATION DEVICE; SYRINGE, BALLOON INFLATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - CORK ENCORE BALLOON CATHETER INFLATION DEVICE; SYRINGE, BALLOON INFLATION Back to Search Results
Model Number M001151050
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Patient Involvement (2645)
Event Date 10/03/2017
Event Type  malfunction  
Manufacturer Narrative
Age at time of event: 18 years or older.(b)(4).
 
Event Description
It was reported that a foreign object was present inside the device syringe.An encore balloon catheter inflation device was selected for use.During unpacking, it was noted that a foreign object of dirt was found inside the syringe.The procedure was completed with another of the same device.No patient complications were reported.
 
Manufacturer Narrative
Device evaluated by mfr: the encore unit was returned with a white stain on the side of the device, the stain was noted to be on the inside of the clear housing and on the outside of the barrel.The stain did not obstruct the graduation marks in the barrel.Excess of silicone in the flexcil line and within the barrel of the device was observed.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The root cause is user preference issue as the product met specification but the user was reportedly dissatisfied with the function, performance, or appearance of the product.(b)(4).
 
Event Description
It was reported that a foreign object was present inside the device syringe.An encore balloon catheter inflation device was selected for use.During unpacking, it was noted that a foreign object of dirt was found inside the syringe.The procedure was completed with another of the same device.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENCORE BALLOON CATHETER INFLATION DEVICE
Type of Device
SYRINGE, BALLOON INFLATION
Manufacturer (Section D)
BOSTON SCIENTIFIC - CORK
business and technology park
model farm road
cork
EI 
Manufacturer (Section G)
BOSTON SCIENTIFIC - CORK
business and technology park
model farm road
cork
EI  
Manufacturer Contact
sonali arangil
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key6963937
MDR Text Key90659142
Report Number2134265-2017-10428
Device Sequence Number1
Product Code MAV
Combination Product (y/n)N
PMA/PMN Number
K955869
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/03/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/19/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/17/2019
Device Model NumberM001151050
Device Catalogue Number15-105
Device Lot Number0020893267
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/10/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/26/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/18/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-