• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS CLINITEK STATUS+ USA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS CLINITEK STATUS+ USA Back to Search Results
Catalog Number 10379675
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Patient Involvement (2645)
Event Date 10/18/2017
Event Type  malfunction  
Manufacturer Narrative
The instrument has been requested to be returned for analysis.No root cause is available at this time.The instrument is currently not being used by the customer.
 
Event Description
The customer reported that their clinitek status+ started smoking when they replaced the batteries.There was no report of injury due to this event.
 
Manufacturer Narrative
The investigation of the returned instrument determined that the cause of the issue was due to a failure of the d16 on the main pcb.D16 is the overvoltage protector diode, which blows if the voltage applied to the instrument is too high.The customer is operational with a replacement instrument.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLINITEK STATUS+ USA
Type of Device
CLINITEK STATUS+
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS
511 benedict avenue
tarrytown NY 10591
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS MANUFACTURING, LTD.
northern road
chilton industrial estate
sudbury, CO10 2XQ
UK   CO10 2XQ
Manufacturer Contact
steven andberg
2 edgewater drive
norwood, MA 02062
7812693655
MDR Report Key6965761
MDR Text Key90223119
Report Number3002637618-2017-00139
Device Sequence Number1
Product Code JIL
UDI-Device Identifier00630414574639
UDI-Public00630414574639
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K091216
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 12/04/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/20/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number10379675
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/15/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-