• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1 Back to Search Results
Model Number 802110
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Patient Involvement (2645)
Event Date 04/05/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The field service representative (fsr) did preventive maintenance (pm) on the unit and the unit operated to the manufacturer's specifications.This complaint is related to (b)(4) / medwatch #1828100-2017-00482.This complaint is related to (b)(4) / medwatch #1828100-2017-00483.
 
Event Description
It was reported that during log analysis of the device, the air bubble detector (abd) was intermittently reporting an "8v power supply error" and "entering broken mode 56" error.There was no patient involvement.
 
Manufacturer Narrative
The reported complaint was confirmed.Data logs were obtained by the field service representative (fsr) while performing preventive maintenance (pm).The logs were reviewed for identified error messages.8v target events were found and referred to the manufacturer's technical (tech) support.Tech support reviewed the logs and determined that this event was related to an 8v target event.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED PERFUSION SYSTEM 1
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
Manufacturer Contact
katie hoyt
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key6966133
MDR Text Key90770083
Report Number1828100-2017-00484
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K153376
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/20/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number802110
Device Catalogue Number802110
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received01/04/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/20/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-