The customer contacted the siemens customer care center (ccc) to report the discordant sodium results.The ccc checked the system remotely.Quality control (qc) was within range.A siemens customer service engineer (cse) was dispatched to the customer's site.The cse performed power flush.The cse replaced all consumables and imt tubing.The cse ran dilution check, which passed.The cse ran qc, which was acceptable.The cause of the discordant sodium results is unknown.The instrument is performing according to the specifications.No further evaluation of this device is required.
|
The initial mdr 2517506-2017-00786 was filed on (b)(6) 2017.Additional information (10/31/2017): a siemens headquarter support center (hsc) specialist reviewed the instrument data, which indicated a fluid flow problem.Process errors and service actions indicated that there was a maintenance need at the time of discrepant results.Average sodium standard deviation above 0.01 mv is an indicator of a poorly grounded integrated multisensor technology (imt) module, which results in imt data unstable errors.Poor sample quality and the presence of gel, fibrin, and cellular material in the sample(s), causing an obstruction, can be a contributing factor to discordant results.The cause of the discordant results is unknown.The instrument is performing according to specifications.No further evaluation of the device is required.
|