The customer stated that they received erroneous results for one patient sample tested for ise indirect na, ci for gen.2 (sodium, chloride) on a cobas 6000 c (501) module - c501.No erroneous results were reported outside of the laboratory.The customer stated that this was the only sample they had an issue with.The sample initially resulted with a sodium value of 125 mmol/l and repeated as 140 mmol/l.The chloride result was initially 116.6 mmol/l and repeated as 98.5 mmol/l.The repeat results were believed to be correct.The patient was not adversely affected.The sodium and chloride electrode lot numbers and expiration dates were asked for, but not provided.Calibration and quality controls performed both before and after the event were successful.The field service engineer could not determine a cause.He replaced the sipper nozzle, pinch tube, and sipper tube.He cleaned the ise area, performed and air purge, performed a reagent prime and ran ise checks 15 times.He performed precision studies.The customer ran calibration and quality controls.
|
All electrodes used on the analyzer were installed on (b)(6) 2017.The lot number of the sodium electrode was b9091 and the lot number of the chloride electrode was r5082.No other assays were affected in the event.The affected sample was initially run in the primary tube, which was processed on a modular pre-analytics system.The customer pulled the primary tube and manually loaded it on the analyzer for repeat testing.A specific root cause could not be determined based on the provided information.Additional information required for the investigation was requested, but not provided.Possible root causes include air in the sample channel, leakage of current coming from the waste, or an old and/or expired reference electrode.No further issues were reported after the service actions.
|