Model Number MMT-1711K
Device Problems Erratic or Intermittent Display (1182); Device Displays Incorrect Message (2591); Power Problem (3010)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/13/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.The insulin pump involved in this event is the 640g insulin infusion pump, which is not marketed in the united states.However, the device is similar to the paradigm real-time insulin infusion pump, which is marketed in the united states.
 
Event Description
It was reported that the insulin pump alarmed post-reset ram crc alarm.The customer¿s blood glucose level was unknown at the time of the incident.It was reported that pump alarmed for power loss and when new battery was inserted pump was still alarming and display was flashing.After this father inserted a new battery and this time pump started up and alarmed for post-reset ram crc alarm.Advised this is due to pump being without power and advised father to reset time and date and rewind pump.Father will call back if any further assistance is required.The insulin pump will not be returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
640G INSULIN PUMP MMT-1711K
Type of Device
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
MDR Report Key6972045
MDR Text Key90779348
Report Number2032227-2017-60089
Device Sequence Number1
Product Code OYC
UDI-Device Identifier00643169554931
UDI-Public(01)00643169554931
Combination Product (y/n)N
Reporter Country CodeFI
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 10/19/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/24/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-1711K
Device Catalogue NumberMMT-1711K
Device Lot NumberHG0QEH9
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received10/13/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-