Engineering evaluation: the device was returned in three separate pieces: the distal shaft with a partially deployed endoprosthesis still mounted, part of the proximal catheter, and the deployment line attached to the deployment knob.It was reported in the additional information that the delivery system was cut and stent wire were damaged due to surgical removal.Approximately 12.7cm of the endoprosthesis was fully expanded.Approximately 8mm of the endoprosthesis was constrained by the inner braided constraining line.The inner braided constraining line was compressed.Approximately 7mm of the endoprosthesis was partially expanded.There were multiple strut rows in the middle of the expanded endoprosthesis that appeared to be detached from the endoprosthesis.The body tape on these strut rows appeared to be displaced.The strut rows at the start of the expanded proximal end of the endoprosthesis appeared to be detached from the endoprosthesis.The body tape on these strut rows appeared to be displaced.The first strut row from the proximal end of the device appeared to be outwardly bent.In the 15mm of the endoprosthesis that was not fully expanded, the nitinol wire was disrupted.There appeared to be 5 single strands of the deployment line at the proximal end of the endoprosthesis attached to the braided constraining line.The single strands had measurements of approximately 25mm, 10mm, 10mm, 4mm, and 3mm.The dual lumen catheter was attached to the transition and a small portion of the distal shaft.The other end of the catheter appeared frayed and torn.The catheter measured to approximately 26.5cm.There were seven pairs of crimping markings on the catheter.There was approximately 212cm of deployment line (including the single fibers) attached to the deployment knob.There were three single fibers at the end of the deployment line and they measured to approximately 60mm, 10mm, and 7mm.There was a single knot in the deployment line approximately 28cm from the deployment knob.Based on the device examination performed, no manufacturing anomalies were identified.Manufacturing plant was corrected.Event description was updated.
|